Glucose Uric Acid Test Device
How To Use Glucose Uric Acid Test Device
Specifications
Blood volume for blood glucose | About 0.6μL |
Blood volume for uric acid | About 3μL |
Sample type | Capillary whole blood, Venous whole blood |
Calibration | Plasma equivalent |
Measuring time for blood glucose | 5±1s |
Measuring time for uric acid | 25±1s |
Meter storage/transportation conditions | -20℃~55℃ |
Dimension | 108*66*22(mm) |
Weight | About 100g |
Power source | 3V DC, 10mA, 2 AAA alkaline batteries |
Battery life | perform up to 1,000 tests |
Display | LCD |
Memory | 500 blood glucose test results with date and time 100 blood glucose control solution test results with date and time100 uric acid test results with date and time,100 uric acid control solution test results with date and time, |
Operating condition for blood glucose | Refer to the user manual of blood glucose test strip. |
Operating condition for uric acid | Refer to the user manual of uric acid test strip. |
Construction | Hand-held |
Measurement units for blood glucose | mg/dL or mmol/L |
Measurement units for uric acid | mg/dL or μmol/L |
Measurement range for blood glucose | 20~600 mg/dL or 1.1~33.3 mmol/L |
Measurement range for uric acid | 3.0~20.0 mg/dL or 181~1188 μmol/L |
Software Version | V01 |
Glucose Test Strip
How To Test Glucose With Strip?
- step1. Insert blood glucose test strip;
- step2. Apply the blood sample;
- step3. Result will be displayed in 5 seconds.
Performance Characteristics
Accuracy: 95% test result of glucose meet the requirement below:
Concentration Range | Bias% |
<5.5 mmol/L(100 mg/dL) | Within ± 0.83 mmol/L(± 15mg/dL) |
≥5.5 mmol/L(100mg/dL) | Within ± 15% |
Precision: test result of glucose meet the requirement below:
Concentration Range | Requirement |
<5.5 mmol/L(100 mg/dL) | SD<0.34mmol/L(6 mg/dL) |
≥5.5 mmol/L(100mg/dL) | CV<6.0% |
Uric Acid Test Strip
How To Test Uric Acid With Strip?
- step1. Insert uric acid test strip;
- step2. Apply the blood sample;
- step3. Result will be displayed in 25 seconds.
Performance Characteristics:
Accuracy: 95% test result of uric acid meet the requirements below:
Concentration Range | Bias% |
≤297 μmol/L(5 mg/dL) | ≤ ± 59.4 μmol/L(1 mg/dL) |
>297 μmol/L(5 mg/dL) | ≤ ± 20% |
Precision: Test result of uric acid meet the requirements below:
Concentration Range | Requirement |
≤297 μmol/L(5 mg/dL) | SD < 22.2 μmol/L(0.37mg/dL) |
>297 μmol/L(5 mg/dL) | CV < 7.5% |
Details please refer to blood glucose and uric acid package insert.
Assessment Report
See Glucose Clinical Evaluation Report here:
See Uric Acid Clinical Evaluation Report here: